Device product
Apollo XT Suture Anchor, 5.5 x 15 Product Usage: Apollo XT Suture Anchors are intended for: Shoulder - Rotator Cuff Repair - Bankart Repair - SLAP Lesion Repair - Biceps Tenodesis - Acromio-Clavicular Separation Repair - Deltoid Repair - Capisular Shift or Capsulolabral Reconstruction Foot/Ankle - Lateral Stabilization - Medial Stabilization - Achilles Tendon Repair Knee - Medial Collateral Ligament Repair - Lateral Collateral Ligament Repair - Posterior Oblique Ligament Repair - Illiotibial Band Tenodesis Elbow - Biceps Tendon Reattachment - Ulnar or Radial Collateral Ligament Reconstruction Hip - Capsular Repair - Acetabular Labral Repair
Z-2430-2018
Product summary
- Event
- Event 80069
- Status
- Terminated
- Classification
- Class II
- Quantity
- 113 units
- Official record key
device-enforcement:Z-2430-2018
Official wording
Reason: The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.
Code information: Model No. XT-5515, Lot No. 2358, 2362, 2363
Distribution pattern: US Nationwide Distribution in the states of California and Oklahoma.
Derived failure modes
-
Unknown
The label on Apollo XT 5.5mm Suture Anchor (part # XTS-5515; lots 2358, 2362, and 2363) indicated that the anchor screw material was titanium, however that part number is made of PEEK.