CS-7, Model No. A4C1

Z-2281-2018

May 09, 2018

Class II

Product summary

Firm
Konica Minolta Medical Imaging USA, Inc.
Event
Event 80220
Status
Terminated
Classification
Class II
Quantity
Unavailable
Official record key
device-enforcement:Z-2281-2018

Official wording

Reason: Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.

Code information: All version numbers.

Distribution pattern: Nationwide distribution. Foreign distribution to Argentina, Brazil, Canada, Chile, Colombia, Ecuador, Japan, Mexico, Netherlands, and Peru.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.