Device product
Elekta Precise Treatment Table
Z-1238-2015
Product summary
- Firm
- Elekta, Inc.
- Event
- Event 70458
- Status
- Terminated
- Classification
- Class II
- Quantity
- 529 US 2951 Worldwide
- Official record key
device-enforcement:Z-1238-2015
Official wording
Reason: It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.
Code information: Precise Treatment Table Serial Numbers: (124001-125803, 125807-880, 125883-899, 125902-917, 125920-949, 125952-963, 125966-1256453, 126158-170, 126174-208, 126212-230, 126232-261, 126269-281, 126293-309,123318-330, 126338-126373, 126375-397, 126400-133999, 213000+)
Distribution pattern: Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland and Syria.
Derived failure modes
-
Unknown
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.