Device product
Vest, Single Item, Product Number Vest, Part Numbers ECREVESTD, ECREVESTF, ECREVESTXLF, REVEST*M, REVESTB, REVESTC, SEVEST*F, SEVEST*M, SEVESTA, SEVESTB, SEVESTC, UEVESTB, UEVESTXLM Product Usage: Products are designed to protect healthcare workers against scatter (secondary) radiation while performing fluoroscopic and radiographic procedures by reducing the amount of radiation reaching the user.
Z-2113-2019
Product summary
- Event
- Event 83144
- Status
- Terminated
- Classification
- Class II
- Quantity
- 22
- Official record key
device-enforcement:Z-2113-2019
Official wording
Reason: The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.
Code information: Serial Numbers: 382976 382964 385618 391130 391143 382858 381860 385617 381858 381854 387903 391141 442855 381789 387954 382861 391150 387944 387965 419791 385623 414166
Distribution pattern: US Nationwide distribution. The products were distributed to the following foreign countries: Korea, Thailand.
Derived failure modes
-
Unknown
The protective material used in the manufacture of the products could potentially provide inadequate protection for the product user. Use of the products could expose the user to scatter radiation of varying degrees depending on many different variables (e.g., distance, shielding, position, time, kV, etc.). Long-term exposure to scatter radiation may increase the risk of cancer, infertility, or other health-related conditions.