Recall Observatory FDA recall evidence

Device product

Olympic Cool-Cap System is intended to provide treatment for neonatal hypoxic-ischemic encephalopathy (HIE). The Cool-Cap is indicated for use in full term infants (more than 36 week gestational age) with clinical evidence of moderate to severe HIE. The device provides selective head cooling with mild systemic hypothermia to prevent or reduce the severity of neurologic injury associated with HIE. It is a selective head cooling system for treatment of neonatal hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in part: "***OLYMPIC COOL-CAP Cooling Unit***Model No.***Made in USA***Software in this product is Natus Medical Incorporated***"

Z-1843-2012

May 21, 2012

Class II

Product summary

Firm
Natus Medical Incorporated
Event
Event 61946
Status
Terminated
Classification
Class II
Quantity
181 units
Official record key
device-enforcement:Z-1843-2012

Official wording

Reason: Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.

Code information: Cooling Serial Numbers: 1008; 1009; 1011; 1012; 1014; 1016; 1017; 1018; 1020; 1021; 1022; 1023; 1024; 1025; 1027; 1028; 1030; 1031; 1034; 1035; 1036; 1038; 1044; 1045; 1048; 1050; 1051; 1052; 1053; 1054; 1055; 1056; 1059; 1060; 1062; 1063; 1064; 1065; 1066; 1070; 1071; 1072; 1073; 1074; 1075; 1076; 1077; 1080; 1081; 1082; 1083; 1085; 1086; 1087; 1088; 1090; 1091; 1092; 1093; 1094; 1095; 1096; 1097; 1101; 1116; 1117; 1118; 1119; 1120; 1122; 1124; 1125; 1126; 1127; 1128; 1129; 1130; 1131; 1132; 1136; 1137; 1138; 1139; 1141; 7002; 7003; 7004; 7005; 7007; 7008; 7009; 7010; 7011; 7013; 7014; 7016; 7017; 7018; 7019; 7020; 7021; 7022; 7023; 7024; 7025; 7026; 7027; 7028; 7030; 7031; 7032; 7033; 7034; 7036; 7037; 7038; 7039; 7040; 7041; 7042; 7043; 7044; 7045; 7046; 7047; 7048; 7049; 7051; 7052; 7053; 7054; 7055; 7250; 7251; 7252; 7253; 7255; 7256; 7257; 7258; 7259; 7260; 7261; 7262; 7263; 7264; 7265; 7266; 7267; 7268; 7269; 7270; 7271; 7272; 7273; 7274; 7275; 7276; 7277; 7278; 7279; 7280; 7281; 7282; 7283; 7284; 7285; 7286; 7288; 7289; 7290; 7291; 7292; 7293; 7294; 7363; 7364; 7365; 7366; 7372; and 7374.

Distribution pattern: Worldwide Distribution -- USA (nationwide) including the states of : AK; AR; AZ; CA; CO; DE; FL; GA; HI; ID; IL; IN; KY; MA; MI; MN; MS; ND; NE; NJ; NY; OH; PA; SC; SD; TN; TX; and WA. and the countries of : Argentina; Australia; Austria; Canada; Egypt; Germany; Hong Kong; Indonesia; Iran; Italy; Kuwait; Poland; Russia; Saudi Arabia; Singapore; Slovenia; South Africa; South Korea; Spain; Taiwan; Thailand; Turkey; United Arab Emirates; and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.