Device product

Devices includes a Multi Port Inflator or without Multi Port Inflator (MPI). All the devices are vascular blood flow machines. Devices are distributed with pressure cuffs of varying sizes as accessories. Affected models include *2100-SX, 3000 series, 2017, 2016, 2015, 2014, 1059, 1059-A, *1059-C, *1058-C, 4000 series, 1100 series, 1060 series, 1080 series, 234 and 246. (* = current production model) All the devices are used in the study and diagnosis of vascular disease and blockages in the upper and lower extremities.

Z-0443-2013

August 08, 2012

Class II

Product summary

Firm
Parks Medical Electronics, Inc
Event
Event 63235
Status
Terminated
Classification
Class II
Quantity
349 units (Domestic: 270 units - Foreign: 79 units)
Official record key
device-enforcement:Z-0443-2013

Official wording

Reason: Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.

Code information: Domestic Distribution: 013053, 013104, 013107, 013124, 013154, 013182, 013193, 014203, 014237, 014248, 014249, 014252, 014253, 014254, 014269, 014276, 014310, 014335, 014339, 014361, 014366, 014376, 014380, 014381, 014382, 014384, 014397, 014401, 014406, 014416, 014424, 014429, 014431, 014443, 014444, 014448, 014454, 014458, 014530, 014532, 014533, 014540, 014542, 014545, 014547, 014551, 014552, 014552, 014553, 014598, 014601, 014604, 014608, 014616, 014654, 014657, 014659, 014660, 014661, 014663, 103077, 105X-600001024, 105X-600001025, 105X-600001026, 105X-600001027, 105X-800001019, 105X-800001022, 105X-800001023, 105X-800001024, 105X-800001026, 105X-800001027, 105X-800001027, 105X-800001028, 105X-800001029, 105X-800001030, 105X-800001031, 105X-E00001001, 105X-E00001002, 105X-E00001003, 105X-E00001004, 105X-E00001005, 105X-E00001007, 105X-E00001008, 105X-E00001009, 105X-E00001010, 105X-E00001011, 105X-E00001012, 105X-E00001013, 105X-E00001014, 105X-E00001016, 105X-E00001017, 105X-E00001018, 105X-E00001019, 105X-E00001020, 105X-F00001000, 105X-F00001001, 105X-F00001002, 105X-F00001003, 105X-F00001004, 105X-F00001005, 105X-F00001005, 105X-F00001006, 105X-F00001007, 105X-F00001008, 105X-F00001009, 105X-F00001010, 105X-F00001011, 105X-F00001012, 105X-F00001013, 105X-F00001014, 105X-F00001015, 105X-F00001016, 105X-F00001017, 105X-F00001018, 105X-F00001020, 105X-F00001021, 105X-F00001022, 105X-F00001023. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify. Foreign distribution: 01301, 013043, 013044, 013045, 013046, 013072, 013076, 013175, 013181, 013183, 013188, 014215, 014386, 014392, 014393, 014456, 014535, 014550, 014551, 014561, 014562, 014563, 105X-800001025, 105X-922001017, 105X-922001018, 105X-922001019, 105X-922001021, 105X-9221020, 105X-923001001, 105X-923001002, 105X-923001003, 105X-923001004, 105X-923001005, 105X-923001006, 105X-923001007, 105X-923001007, 105X-923001009, 105X-923001010, 105X-923001011, 105X-E00001000, 105X-E00001003, 105X-E00001006, 105X-E00001015, 105X-F00001019. There are also devices that pre-dated bar codes, as well as devices with unknown codes that the firm is unable to identify.

Distribution pattern: Worldwide distribution: USA (nationwide) and countries of: Australia, Brazil, Canada, Chile, Egypt, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Korea, Mexico, Peru, Republica Argentina, Spain, Switzerland, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Parks Medical Electronics, Inc. was made aware of two incidents where an air supply tube from their PARKS PVR unit was inadvertently connected to an IV in the patient's arm instead of into the blood pressure cuff.