Device product

The LIFEPAK 500 Automated External Defibrillator is a semi-automatic defibrillator that uses a patented Shock Advisory System. This software algorithm analyzes the patient's electrocardiographic (ECG) rhythm and indicates whether or not it detects a shockable rhythm. The LIFEPAK 500 AED is intended for use by personnel who are authorized by a physician/medical director and is intended for use in the pre-hospital and workplace and community environments.

Z-1773-2013

June 12, 2013

Class II

Product summary

Firm
Physio Control, Inc.
Event
Event 65565
Status
Terminated
Classification
Class II
Quantity
139 total (125 units in the US and 14 units outside US)
Official record key
device-enforcement:Z-1773-2013

Official wording

Reason: The LIFEPAK 500 Automated External Defibrillator (AED) could be susceptible to malfunction due to an assembly error. If affected, the malfunction would be the result of internal shorting on a printed circuit board assembly and could render the device inoperable.

Code information: PART NUMBERS: 3011790-001568, 3011790-001574, 3011790-001627, 3011790-001634, and 3011790-001650. SERIAL NUMBERS OUTSIDE the US: 41232265, 41232267, 41232269, 41232270, 41232271, 41232272, 41238207, 41238208, 41301145, 41301370, 41301371, 41301372, 41301373, and 41301374. SERIAL NUMBERS IN THE US: 41134905, 41134906, 41134907, 41135982, 41135983, 41135984, 41135991, 41135992, 41135993, 41135994, 41135995, 41135996, 41135997, 41135998, 41135999, 41136000, 41136001, 41136002, 41136003, 41136004, 41136005, 41136006, 41136007, 41136008, 41136009, 41136010, 41136011, 41136012, 41136013, 41136014, 41136015, 41136016, 41136017, 41136018, 41136019, 41136020, 41136021, 41136022, 41136023, 41136024, 41136025, 41136026, 41136027, 41136028, 41136029, 41136030, 41136031, 41136033, 41136034, 41136035, 41136036, 41136037, 41136038, 41136039, 41136040, 41136060, 41136061, 41136062, 41136063, 41136064, 41136065, 41136066, 41136067, 41136068, 41136069, 41136070, 41136071, 41136072, 41136073, 41136074, 41136075, 41136076, 41136077, 41136078, 41136079, 41136092, 41136093, 41136094, 41142873, 41145359, 41145360, 41145361, 41145362, 41145363, 41145364, 41145365, 41145366, 41145367, 41145368, 41145369, 41145370, 41145371, 41145372, 41145373, 41145374, 41145375, 41145376, 41145377, 41145378, 41164115, 41232002, 41232003, 41232004, 41232005, 41232249, 41232250, 41232251, 41232252, 41232253, 41232254, 41232255, 41232256, 41232257, 41232258, 41232259, 41232260, 41232554, 41239182, 41239183, 41239184, 41239185, 41239186, 41239187, 41239188, and 41239189.

Distribution pattern: Worldwide distribution: US (nationwide) including states of: IL, MA, NC, ND, NY, PA, RI, TX, WA, and WI; and country of: Canada.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    assembly error