Skip to content
Recall Observatory FDA recall evidence

Device product

ETHICON SURGIFLO HEMOSTATIC MATRIX Distributed by ETHICON ETHICON, Inc. Somerville, NJ 08876, USA SURGIFLO Hemostatic Matrix is a topical absorbable hemostat consisting of a flowable gelatin matrix (paste). This kit also contains components necessary for preparing and applying the matrix during surgery. SURGIFLO Hemostatic Matrix, mixed with sterile saline or thrombin solution, is indicated in surgical procedures (other than ophthalmic) as an adjunct to hemostasis when control of bleeding by ligature or other conventional methods is ineffective or impractical.

Z-0293-2013

August 21, 2012

Class I

Product summary

Firm
Ethicon, Inc.
Event
Event 63322
Status
Terminated
Classification
Class I
Quantity
45,200
Official record key
device-enforcement:Z-0293-2013

Official wording

Reason: SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.

Code information: PMA number P990004 Medical Device Listing Number D100622 Product Code: 2991 Lot # 239091 Exp Mar-13 Lot # 238783 Exp Sep-13 Lot # 238635 Exp Mar-13 Lot # 237498 Exp Mar-13 Lot # 237119 Exp Mar-13 Lot # 235488 Exp Jan-13

Distribution pattern: Worldwide Distribution - USA (nationwide) and the countries of Russia, China, and the European Union.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    SURGIFLO Hemostatic Matrix Kit & SURGIFLO Hemostatic Matrix Kit with Thrombin have an issue within the packing process where a cut could potentially breach the double Tyvek pouch of the packaging.