Device product
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Z-1370-2015
Product summary
- Event
- Event 70678
- Status
- Terminated
- Classification
- Class II
- Quantity
- 72 devices
- Official record key
device-enforcement:Z-1370-2015
Official wording
Reason: The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Code information: Serial Nos. E34297300656, E34297300657, E34297300659, E34297300660, E34297300661, E34297300666, E34297300667, E34297300668, E34297300669, E34297300670, E34297300671, E34297300673, E34297300675, E34297300676, E34297300677, E34297300678 E34297502563, E34297502564, E34297502565, E34297502566, E34297502567, E34297502568, E34297502569, E34297502570, E34297502571, E34297502572, E34297502573, E34297502574, E34297502575, E34297502576, E34297502577, E34297502578, E34297502579, E34297502580, E34297502581, E34297502583, E34297502584, E34297502585, E34297502586, E34297502587, E34297502588, E34297502589, E34297502590, E34297502591, E34297502592, E34297502593, E34297502594, E34297502595, E34297502596, E34297502597, E34297502599, E34297502600, E34297502601, E34297502602, E34297502603, E34297502604, E34297502605, E34297502606, E34297502607, E34297502608, E34297502612, E34297502613, E34297502619, E34297502620, E34297502621, E34297502629, E34297600499, E34297600501, E34297600502, E34297600503, E34297600507, E34297600508.
Distribution pattern: Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim
Derived failure modes
-
Unknown
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.