Device product
Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.
Z-0627-2016
Product summary
- Firm
- Bayer Healthcare
- Event
- Event 72817
- Status
- Terminated
- Classification
- Class II
- Quantity
- 198,598
- Official record key
device-enforcement:Z-0627-2016
Official wording
Reason: The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail
Code information: 50856405, 50857510, 50860994, 50866789, 50867921, 50869464, 50869465, 50871131, 50872965, 50878229, 50878230, 50878231, 50878232, 50881328, 50881393, 50881394, 50881395, 50881506, 50881511, 50881512, 50882753, 50886316, 50886467, 50886468, 50886469, 50888285, 50888286, 50888287, 50888288, 50888373, 50890483, 50890484, 50890485, 50892357, 50892358, 50892359, 50892360, 60000321, 60000322, 60000323, 60000692, 60000789, 60000790, 60000833, 60000834, 60000835
Distribution pattern: Nationwide Distribution to the states of : AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT and WI.
Derived failure modes
-
Unknown
The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail