Device product

Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.

Z-0627-2016

November 30, 2015

Class II

Product summary

Firm
Bayer Healthcare
Event
Event 72817
Status
Terminated
Classification
Class II
Quantity
198,598
Official record key
device-enforcement:Z-0627-2016

Official wording

Reason: The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail

Code information: 50856405, 50857510, 50860994, 50866789, 50867921, 50869464, 50869465, 50871131, 50872965, 50878229, 50878230, 50878231, 50878232, 50881328, 50881393, 50881394, 50881395, 50881506, 50881511, 50881512, 50882753, 50886316, 50886467, 50886468, 50886469, 50888285, 50888286, 50888287, 50888288, 50888373, 50890483, 50890484, 50890485, 50892357, 50892358, 50892359, 50892360, 60000321, 60000322, 60000323, 60000692, 60000789, 60000790, 60000833, 60000834, 60000835

Distribution pattern: Nationwide Distribution to the states of : AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT and WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail