Device product

Female Driver, Part number FDr101, instrument used for the insertion of the Fassier-Duval IM Telescopic System. For orthopedic procedures. This instrument is included in the Fassier-Duval IM Telescopic System instrument tray. This instrument is marked with the catalogue number and lot number.

Z-1546-2015

April 10, 2015

Class II

Product summary

Firm
Pega Medical Inc.
Event
Event 71061
Status
Terminated
Classification
Class II
Quantity
9 units (5 units in the US and 4 units international)
Official record key
device-enforcement:Z-1546-2015

Official wording

Reason: The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.

Code information: Lot 290609-01

Distribution pattern: Worldwide distribution. US in Delaware, Florida, Nebraska, and New York; Australia, Brazil, and United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The hexagonal tip of the Female Driver instrument for the Fassier-Duval IM Telescopic System is 0.06 mm under the specifications. This may result in a reduction of the total strength of the instrument, possibly leading to deformation, wear, and tip breakage.