Device product
Beacon Tip Royal Flush Plus High-Flow Catheter. Angiographic catheter. For use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.
Z-2318-2015
Product summary
- Firm
- Cook Inc.
- Event
- Event 71692
- Status
- Terminated
- Classification
- Class I
- Quantity
- 268169
- Official record key
device-enforcement:Z-2318-2015
Official wording
Reason: Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.
Code information: All lots of the following Product numbers: HNR4.0-35-100-P-10S-0, HNR4.0-35-100-P-10S-CFP, HNR4.0-35-100-P-10S-PIG, HNR4.0-35-100-P-6S-VCF, HNR4.0-35-100-P-NS-0, HNR4.0-35-100-P-NS-DAV, HNR4.0-35-100-P-NS-SIM1, HNR4.0-35-100-P-NS-SIM2, HNR4.0-35-100-ST-10S-0, HNR4.0-35-100-ST-10S-PIG, HNR4.0-35-100-ST-6S-PED, HNR4.0-35-110-P-10S-PIG, HNR4.0-35-120-P-6S-PED, HNR4.0-35-120-ST-6S-PED, HNR4.0-35-125-P-10S-0, HNR4.0-35-125-P-10S-PIG, HNR4.0-35-50-P-NS-0, HNR4.0-35-65-P-8S-VCF, HNR4.0-35-70-P-10S-0, HNR4.0-35-70-P-10S-CFP, HNR4.0-35-70-P-10S-PIG, HNR4.0-35-70-P-4S-PED, HNR4.0-35-70-P-6S-PED, HNR4.0-35-70-P-NS-0, HNR4.0-35-70-ST-10S-0, HNR4.0-35-70-ST-10S-PIG, HNR4.0-35-70-ST-6S-PED, HNR4.0-35-90-P-10S-0, HNR4.0-35-90-P-10S-CFP, HNR4.0-35-90-P-10S-PIG, HNR4.0-35-90-P-6S-PED, HNR4.0-35-90-P-8S-VCF, HNR4.0-35-90-P-NS-0, HNR4.0-35-90-ST-10S-0, HNR4.0-35-90-ST-10S-PIG
Distribution pattern: Worldwide Distribution - US (nationwide including Puerto Rico and US Virgin Islands), Algeria, Antigua, Armenia, Austria, Australia, Bahamas, Belgium, Bangladesh, Bulgaria, Bahrain, Brazil, Brunei Darussalam, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Ghana, Great Britain, Greece Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kuwait, Kazakhstan, Republic of Korea, Kuwait, Lebanon, Lithuania, Luxemburg, Latvia, Libya, Macau, Malaysia, Malta, Martinique, Mexico, Mongolia, Morocco, Netherlands, Netherlands Antilles, New Caledonia, New Zealand, Nicaragua, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Sri Lanka, Sweden, Switzerland, Spain, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Derived failure modes
-
Unknown
Reports of catheter tip splits and or separation. Loss of device function, medical intervention to retrieve separated segment, or complications resulting from separated tip occluding blood flow to end organs may result.