Skip to content
Recall Observatory FDA recall evidence

Device product

EXACTECH Equinoxe Narrow Modular Reverse Pilot-Tip Reamer // Catalog #: 321-45-38, Catalog #: 321-45-42 and Catalog #: 321-45-46

Z-2668-2017

May 25, 2017

Class II

Product summary

Firm
Exactech, Inc.
Event
Event 77004
Status
Terminated
Classification
Class II
Quantity
198
Official record key
device-enforcement:Z-2668-2017

Official wording

Reason: Instructions for use (IFU) is updated to clarify the proper technique and instrumentation to ream the Glenoid, i.e., to include a caution statement that the surgeon should ream on-axis and that reaming off-axis may result in instrument breakage.

Code information: All Lots

Distribution pattern: AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, Hawaii and Puerto Rico Australia, Austria, Brazil, Canada, China, France, Germany, Greece, India, Italy, Korea, Luxembourg, Singapore, South Africa, Spain, Sweden, Switzerland, The Netherlands and United Kingdom

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0