ACL TOP 300 and ACL TOP 300 CTS (Japan); PN 0000280060 Bench top, fully automated, random access analyzer designed specifically for in vitro diagnostic clinical use in the hemostasis laboratory for coagulation and/or fibrinolysis testing in the assessment of thrombosis and/or hemostasis. The systems provide results for both direct hemostasis measurements and calculated parameters.

Z-2249-2013

July 19, 2013

Class II

Product summary

Firm
Instrumentation Laboratory Co.
Event
Event 65893
Status
Terminated
Classification
Class II
Quantity
662
Official record key
device-enforcement:Z-2249-2013

Official wording

Reason: Potential carryover issue that can can cause shortened APTT clotting times on the instrument.

Code information: Start SN and Date: 12060256, Jun 2012; End SN and Date: 13070837, July 2013 and Start SN and Date: 11120131J, Dec 2011; End SN and Date: 13060320J, Jun2013

Distribution pattern: Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. Internationally to Canada, Argentina, Australia, Austria, Belarus, Belgium, Brazil, Brunei, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dubai, Finland, France, Germany, Great Britian, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Luthuania, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Uruguay, Venezuela, and Viet Nam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential carryover issue that can can cause shortened APTT clotting times on the instrument.