Device product

ADVIA Centaur XPT Immunoassay System

Z-0072-2017

April 28, 2016

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 74162
Status
Terminated
Classification
Class II
Quantity
292 units Total (3 domestically & 289 internationally)
Official record key
device-enforcement:Z-0072-2017

Official wording

Reason: The ADVIA Centaur¿ XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.

Code information: None

Distribution pattern: Worldwide Distribution - US including MD, MN, & MS; **Internationally** AE, AT, AU, BE, BR, CA, CN, DE, DK, EG, ES, FI, FR, GB, GP, HU, IN, IT, JO, JP, KR, MY, NO, NZ, PH, PT, QA, SA, SE, SG, TR & VN

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The ADVIA Centaur¿ XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.