Device product
Syngo.plaza, Picture Archiving and Communications System(PACS), Model 10863171 Syngo.plaza is a Picture Archiving and Communication System (PACS) intended to display, process, read, report, communicate, distribute, store and archive digital medical images. It supports the physician in diagnosis and treatment planning. Syngo.plaza also supports storage and archiving of DICOM Structured reports. In a comprehensive imaging syngo.plaza integrates Hospital/Radiology Information Systems (HIS/RIS) to enable customer specific work flows. Syngo.plaza optionally uses a variety of advanced postprocessing applications.
Z-0168-2018
Product summary
- Event
- Event 77456
- Status
- Terminated
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-0168-2018
Official wording
Reason: Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.
Code information: non-medical software application Check Tool V1 .0 (L TA Incomplete Archive Check Tool V1 .0)
Distribution pattern: Distribution to MN
Derived failure modes
-
Unknown
Siemens is releasing a non-medical software application LTA Incomplete Archive Check Tool V1.0. It is intended to eliminate an issue that occurs during syngo.plaza de-archiving from Dicom LTA. It has been defined that the number of de-archived images is less than the count of the archived images for the series. The issues were discovered through Siemens Global Complaint Handling System.