Device product

Medtronic CareLink iPro Therapy Management Software, MMT-7340 The CareLink iPro Therapy Management Software (model, MMT-7340) is intended to be used by a healthcare professional (HCP) with the iPro2 CGM System and provides upload support for the iPro2 Recorder. The CareLink iPro software is used to upload glucose sensor data stored on a recording device, capture BG meter readings through manual entry or wired-upload, retrospectively calibrate sensor data, and provide retrospective reports of continuous glucose information. The reports generated by the CareLink iPro software are intended for use by healthcare professionals to assist with diabetes clinical assessment, patient education, and therapy optimization.

Z-0670-2018

October 10, 2017

Class II

Product summary

Firm
Medtronic Inc.
Event
Event 78298
Status
Terminated
Classification
Class II
Quantity
488 uploads
Official record key
device-enforcement:Z-0670-2018

Official wording

Reason: Due to a time conversion error, data uploaded from the iPro(TM)2 recorder to the CareLink iPro software using the new CareLink iPro Uploader feature may result in incorrect reports. This pertains only to reports generated between September 23, 2017 and September 27, 2017 that used the new CareLink iPro Uploader feature.

Code information: Version 1.12

Distribution pattern: Worldwide Distribution - US Distribution

Derived failure modes

  • Device software or design

    reason.device_software_design · v1.0.0

    software using the new CareLink iPro Uploader feature may result in incorrect