Elekta Unity
Z-0706-2019
Product summary
- Firm
- Elekta Limited
- Event
- Event 81886
- Status
- Completed
- Classification
- Class II
- Quantity
- 2 units
- Official record key
device-enforcement:Z-0706-2019
Official wording
Reason: In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.
Code information: Serial Numbers: 1536549, 1536549
Distribution pattern: TX and WI, Germany, Denmark, Netherlands, United Kingdom
Derived failure modes
-
Unknown
In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.