Guardian Connect App CSS7200 used on the iPhone, iPad, and iPod Touch devices.

Z-1282-2019

July 04, 2018

Class II

Product summary

Firm
Medtronic Inc.
Event
Event 81656
Status
Terminated
Classification
Class II
Quantity
Approximately 9,328 registered users
Official record key
device-enforcement:Z-1282-2019

Official wording

Reason: The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.

Code information: UDI - (01)00643169693142(10)

Distribution pattern: Distribution was only made OUS. Countries involved are: Austria, United Arab Emirate, Australia, Belgium, Switzerland, Chile, Colombia, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Iceland, Italy, Republic of Korea, Luxembourg, Netherlands, Norway, Poland, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Turkey, and South Africa.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The application may be closed by the operating system without alerting the user the app is no longer running or communicating with the transmitter resulting in the user not receiving alerts that could be associated with hypoglycemic or hyperglycemic events.