Skip to content
Recall Observatory FDA recall evidence

Device product

AutoMate 1250, Sorter with Aliquot Module - UDI: 15099590364106; Model #ODL25125 Automate 2550, High Speed Sorter with Aliquot Module - UDI: 15099590364083 , Model #ODL25255

Z-0492-2020

November 09, 2018

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 83732
Status
Ongoing
Classification
Class II
Quantity
83 automate systems
Official record key
device-enforcement:Z-0492-2020

Official wording

Reason: Beckman Coulter has identified a case where a customer using the microltiterplate kit option without a sorting drive encountered a cross-contamination of samples when a connection time-out occurred between the system and Laboratory Information System (LIS) .

Code information: All units containing software Version(s) 4.1.1 and up

Distribution pattern: US: NY, VA OUS: Austria Belgium Czech Republic France Germany Israel Italy Lithuania Mayotte Romania Slovakia South Africa Spain Sweden Switzerland United Kingdom

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0