Device product

NIDEK Specular Microscope, Model CEM-530 - Product Usage: The device provides noncontact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images.

Z-0771-2020

February 25, 2019

Class III

Product summary

Firm
Nidek, Inc.
Event
Event 84112
Status
Terminated
Classification
Class III
Quantity
13 devices
Official record key
device-enforcement:Z-0771-2020

Official wording

Reason: The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)

Code information: Serial numbers 120080, 120081, 120086, 120087, 120088, 120090, 120091, 120092, 120093, 120094, 120095, 120096, and 170001.

Distribution pattern: US Nationwide distribution in the states of AZ, CA, IL, KS, MD, NJ, NY, PA, SC, and VA, as well as PR.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2)