natus newborn care Olympic Brainz Monitor, REF OBM00002. electroencephalograph
Z-1630-2020
Product summary
- Event
- Event 85189
- Status
- Terminated
- Classification
- Class II
- Quantity
- 40 units
- Official record key
device-enforcement:Z-1630-2020
Official wording
Reason: The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.
Code information: Serial Numbers: OBM00002G2119, OBM00002G2130, OBM00002G2131, OBM00002G2132, OBM00002G2133, OBM00002G2134, OBM00002G2135, OBM00002G2136, OBM00002G2137, OBM00002G2138, OBM00002G2139, OBM00002G2140, OBM00002G2141, OBM00002G2142, OBM00002G2143, OBM00002G2144, OBM00002G2145, OBM00002G2146, OBM00002G2147, OBM00002G2148, OBM00002G2149, OBM00002G2150, OBM00002G2151, OBM00002G2152, OBM00002G2153, OBM00002G2154, OBM00002G2155, OBM00002G2156, OBM00002G2157, OBM00002G2158, OBM00002G2159, OBM00002G2160, OBM00002G2161, OBM00002G2162, OBM00002G2163, OBM00002G2164, OBM00002G2165, OBM00002G2166, OBM00002G2167, OBM00002G2168
Distribution pattern: Worldwide distribution. US nationwide, Australia, Dominican Republic, United Arab Emirates, Israel, Italy, Hong Kong, Korea, Hungary, Switzerland, Colombia, and United Kingdom
Derived failure modes
-
Unknown
The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest to the system functionality of the affected items.