Device product

Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC system is indicated for the incision/cutting removal, drilling and sawing of soft and hard tissue and bone in surgical procedures.

Z-2673-2020

June 12, 2020

Class II

Product summary

Firm
Medtronic Xomed, Inc.
Event
Event 85886
Status
Terminated
Classification
Class II
Quantity
673 units
Official record key
device-enforcement:Z-2673-2020

Official wording

Reason: During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.

Code information: Model: 1898072 Affected Lot Numbers: 0217355465, 0218377490, 0218687535, 0218928638, 0219310085,0219639840, 0217856852, 0218472653, 0218714701, 0218958034, 0219443322,0219639842, 0217858947,0218472654, 0218721203, 0218958035, 0219448447, 0219639844, 0217859045,0218665536, 0218745073, 0219004569,0219465587, 0218203285, 0218666231, 0218746389,0219201122,0219518520,0218263964, 0218666243, 0218864105, 0219255151, 0219550550, 0218293282, 0218687534, 0218900337,0219281168, 0219619754.

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.