Software 1898072 IPC upgrade v 2.7.3.0 - Product Usage: The IPC system is indicated for the incision/cutting removal, drilling and sawing of soft and hard tissue and bone in surgical procedures.
Z-2673-2020
Product summary
- Event
- Event 85886
- Status
- Terminated
- Classification
- Class II
- Quantity
- 673 units
- Official record key
device-enforcement:Z-2673-2020
Official wording
Reason: During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.
Code information: Model: 1898072 Affected Lot Numbers: 0217355465, 0218377490, 0218687535, 0218928638, 0219310085,0219639840, 0217856852, 0218472653, 0218714701, 0218958034, 0219443322,0219639842, 0217858947,0218472654, 0218721203, 0218958035, 0219448447, 0219639844, 0217859045,0218665536, 0218745073, 0219004569,0219465587, 0218203285, 0218666231, 0218746389,0219201122,0219518520,0218263964, 0218666243, 0218864105, 0219255151, 0219550550, 0218293282, 0218687534, 0218900337,0219281168, 0219619754.
Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AL, AK, AZ, AR, CA,CO, CT, FL, GA, HI, ID,IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NY, NC, OH, OK, PA, RI, SC, TN, TX, UT, VA, WV, WA, WI. The countries of Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Greece, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Poland, Singapore, Spain, Sweden, Switzerland, United Kingdom.
Derived failure modes
-
Unknown
During internal testing execution of the next generation of Integrated Power Console (IPC) prototype it was noted that the M5 Microdebrider was rotating at a higher speed than the set value in the cosole.