Device product
SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825
Z-0399-2022
Product summary
- Event
- Event 89035
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1433
- Official record key
device-enforcement:Z-0399-2022
Official wording
Reason: Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.
Code information: Lot: M106450, UDI: 15099590575113
Distribution pattern: US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand
Derived failure modes
-
Foreign material or chemical contamination
particulates