Skip to content
Recall Observatory FDA recall evidence

Device product

CyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Z-1272-2021

February 26, 2021

Class II

Product summary

Firm
Accuray Incorporated
Event
Event 87439
Status
Terminated
Classification
Class II
Quantity
243 units
Official record key
device-enforcement:Z-1272-2021

Official wording

Reason: The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.

Code information: Product Codes: 019200, 020600, 021831, 022986, 023730, 025000, 027600, 028000, 030000, 031000, 031500, 0660000, 029640-001, 029828-003, 031000-004, 032000-001, 032000-004, 033000-000, 033000-010, 053101-010, 053101-M-000, 053301-010, 053301-M-000, 053401-010, 053401-M-000, 054000-002, 054000-004, 0660000-000, 1010140-000, 1047641-000 Serial Numbers: C0063, C0067, C0008, C0076, C0077, C0085, C0098, C0101, C0112, C0126, C0131, C0135, C0138, C0151, C0155, C0162, C0166, C0172, C0174, C0176, C0177, C0186, C0187, C0188, C0195, C0197, C0200, C0206, C0208, C0211, C0213, C0228, C0231, C0234, C0240, C0242, C0246, C0253, C0263, C0266, C0271, C0239, C0441, C0446, C0447, C0456, C0460, C0464, C0465, C0469, C0472, C0475, C0481, C0506, C0507, C0193, C0241, C0293, C0090, C0092, C0093, C0100, C0103, C0104, C0115, C0117, C0119, C0124, C0130, C0148, C0150, C0153, C0168, C0181, C0182, C0194, C0205, C0226, C0235, C0236, C0244, C0254, C0255, C0258, C0259, C0267, C0268, C0274, C0276, C0280, C0282, C0283, C0285, C0286, C0289, C0291, C0292, C0297, C0298, C0299, C0300, C0301, C0302, C0303, C0304, C0305, C0306, C0309, C0310, C0312, C0313, C0317, C0319, C0321, C0325, C0328, C0329, C0331, C0339, C0347, C0353, C0357, C0365, C0377, C0382, C0386, C0388, C0399, C0401, C0403, C0405, C0416, C0432, C0435, C0451, C0064, C0071, C0074, C0091, C0099, C0102, C0137, C0238, C0249, C0252, C0270, C0277, C0294, C0417, C0476, C0384, C0414, C0424, C0439, C0427, C0428, C0344, C0423, C0372, C0387, C0330, C0343, C0345, C0346, C0348, C0351, C0352, C0356, C0360, C0363, C0368, C0369, C0370, C0373, C0380, C0381, C0383, C0389, C0390, C0394, C0395, C0396, C0397, C0398, C0402, C0404, C0406, C0411, C0413, C0415, C0420, C0422, C0429, C0430, C0436, C0437, C0438, C0440, C0444, C0445, C0448, C0453, C0457, C0458, C0466, C0461, C0467, C0473, C0480, C0482, C0484, C0486, C0487, C0488, C0489, C0490, C0491, C0492, C0493, C0494, C0495, C0496, C0497, C0500, C0505, C0513, C0014, C0022, C0040, C0042, C0046, C0047, C0056, C0059, C0066, C0079, C0086, C0089, C0140, C0173, C0311, C0355, C0470

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of AK, AZ. CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, PA, TN, TX, VA, WA, WI, WV, PR and the countries of Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italy, Japan, Republic of Korea, Latvia, Luxembourg, Malaysia, Mexico, Oman, Pakistan, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The set screws that connect the Standard Treatment Couch linkage arm to the roll motor can loosen over time allowing the couch to roll.