Device product
Trilogy Evo O2, Product numbers DE2100X13B, DS2100X11B, EE2100X15B, ES2100X15B, EU2100X15B, FR2100X14B, FX2100X15B, IA2100X15B, IN2100X15B, IN2100X19, IT2100X21B, JP2100X16B, LA2100X15B, ND2100X15B, RDE2100X13B, SP2100X26B
Z-2288-2021
Product summary
- Event
- Event 88333
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 101 (FEMA Stockpile Only); 21077 (OUS)
- Official record key
device-enforcement:Z-2288-2021
Official wording
Reason: Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).
Code information: Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02
Distribution pattern: Worldwide distribution - US nationwide distribution.
Derived failure modes
-
Unknown
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).