Device product
RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs
Z-1598-2021
Product summary
- Event
- Event 87868
- Status
- Terminated
- Classification
- Class II
- Quantity
- N/A
- Official record key
device-enforcement:Z-1598-2021
Official wording
Reason: When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.
Code information: Software version number: 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 , 10.1.0.613
Distribution pattern: Domestic: TN
Derived failure modes
-
Unknown
When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.