Skip to content
Recall Observatory FDA recall evidence

Device product

ExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST; 20910-01A, ETD Positioning and Moni. 1.0.1 INST; 20910-01B, ETD Positioning and Moni. 1.0.2 INST; and 20910-01C ETD Positioning and Moni. 1.0.3 INST.

Z-0733-2022

February 25, 2022

Class II

Product summary

Firm
Brainlab AG
Event
Event 89662
Status
Ongoing
Classification
Class II
Quantity
205 systems
Official record key
device-enforcement:Z-0733-2022

Official wording

Reason: The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.

Code information: Software versions 1.0.0, 1.0.1, 1.0.2, 1.0.3, GTIN 04056481142315.

Distribution pattern: Distribution was made to AZ, AR, CA, CO, FL, GA, IN, KY, MD, MS, MO, NJ, NY, NC, OK, OR, PA, TX, VA, and WA. There was no government or military distribution. Foreign distribution was made to Australia, Austria, Bahrain, Belgium, China, Denmark, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, New Zealand, Poland, Singapore, Slovakia, South Korea, Spain, Thailand, Turkey, United Kingdom, and United Arab Emirates.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.