Skip to content
Recall Observatory FDA recall evidence

Device product

ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1.2.0 or earlier

Z-1292-2022

April 21, 2022

Class II

Product summary

Firm
Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.)
Event
Event 90165
Status
Terminated
Classification
Class II
Quantity
41 devices (out of 1,260 users)
Official record key
device-enforcement:Z-1292-2022

Official wording

Reason: Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.

Code information: prouP Version - 1.2.0 UDI Code: (01)00860006184902(10)v1.x.x

Distribution pattern: Worldwide distribution - US Nationwide. The software app is distributed via the Apple App Store.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0