Device product
LUMINOS Lotus Max (VF11)
Z-1481-2023
Product summary
- Event
- Event 92229
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 704 units in the United States (1945 units worldwide)
- Official record key
device-enforcement:Z-1481-2023
Official wording
Reason: Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.
Code information: Model: 11574100
Distribution pattern: US Nationwide - Worldwide Distribution
Derived failure modes
-
Unknown
Under certain unlikely circumstances during a fluoroscopic examination, the imaging system Fluorospot Compact might sporadically display an incorrect air kerma/air kerma rate related to the patient reference point, according to IEC 60601-2-43. There is no impact on workflow or diagnosis. The error can occur only in fluoroscopy systems with a second (overhead) X-ray tube.