Skip to content
Recall Observatory FDA recall evidence

Device product

Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD

Z-0116-2022

September 17, 2021

Class II

Product summary

Firm
Elekta, Inc.
Event
Event 88681
Status
Terminated
Classification
Class II
Quantity
4,405 units
Official record key
device-enforcement:Z-0116-2022

Official wording

Reason: There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

Code information: a. Elekta Precise Digital Accelerator - UDI GTIN Number: 5060191071017, Model Numbers: XRT 0601 (NFB), XRT 0611 (FB), All Serial Numbers; b. Elekta Harmony - UDI GTIN Number: 05060191071581, Model Numbers: XRT 2201 (NFB), All Serial Numbers; c. Elekta Synergy - UDI GTIN Number: 05060191071550, Model Numbers: XRT 0621 (NFB), XRT 0631 (FB), All Serial Numbers; d. Elekta Infinity - UDI GTIN Number: 05060191071543, 05060191071512, Model Numbers: XRT 0511 (NFB & MLCi2), XRT 0521 (FB & MLCi2), XRT 0531 (NFB & Agility), XRT 0541 (FB & Agility), All Serial Numbers; e. Versa HD - UDI GTIN Number: 05060191071574, 05060191071529, Model Numbers: XRT 2121 (NFB), XRT 2131 (FB), All Serial Numbers

Distribution pattern: Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA, OR, KY, VA, TX, SD, AZ, NJ, MI, WI, ND, MO, WA, OK, PR, NM, NC, ID, CT, DE, LA, MT, TN, CO, UT, MN, IN, RI, IA, DC, ND, AZ, HI, KS, NE, NH, AR, WY, WV, MS, VT and internationally to Albania, Antigua/Barbuda, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kyrgystan, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.