Device product
Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application
Z-0319-2021
Product summary
- Event
- Event 86534
- Status
- Terminated
- Classification
- Class II
- Quantity
- 8 devices
- Official record key
device-enforcement:Z-0319-2021
Official wording
Reason: There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.
Code information: Model 3892 v1.04 is installed on Model 3300 LATITUDE Programming Systems with the following serial numbers: 009494 012689 010356 008206 008124 007983 004813 010751
Distribution pattern: Distributed in US - AR, FL, IL, and TX
Derived failure modes
-
Unknown
There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.