Device product
Thermo SCIENTIFIC, Sensititre CMC5VGNF, Gram Negative IVD AST
Z-0331-2023
Product summary
- Firm
- Remel Inc
- Event
- Event 91104
- Status
- Terminated
- Classification
- Class I
- Quantity
- 588 units
- Official record key
device-enforcement:Z-0331-2023
Official wording
Reason: An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.
Code information: Lot Numbers: B2125, B2115
Distribution pattern: US Nationwide Distribution
Derived failure modes
-
Unknown
An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.