Atellica CI Analyzer. Catalog Numbers: 10947347.

Z-0597-2025

November 04, 2024

Class II

Product summary

Firm
Siemens Healthcare Diagnostics, Inc.
Event
Event 95644
Status
Ongoing
Classification
Class II
Quantity
559 units
Official record key
device-enforcement:Z-0597-2025

Official wording

Reason: Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.

Code information: UDI-DI Number: 630414229560; Catalog Number: 10947347. Software version 1.28.51

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.