Device product

Illumina NextSeq 550Dx, REF: 20005715, CE, IVD

Z-0854-2022

February 22, 2022

Class II

Product summary

Firm
Illumina, Inc.
Event
Event 89699
Status
Completed
Classification
Class II
Quantity
674 instruments
Official record key
device-enforcement:Z-0854-2022

Official wording

Reason: There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.

Code information: Lot #s NDX550110 - NDX550843 ; UDI: (01)00816270020125

Distribution pattern: US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN OUS: Australia Austria Belgium Chile China Denmark Finland France Germany Hungary Ireland Italy Japan Latvia Netherlands Norway Poland Portugal Romania Russian Federation Saudi Arabia Singapore Slovakia South Africa South Korea Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom Vietnam

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.