MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Z-1572-2026

February 02, 2026

Class II

Product summary

Firm
Howmedica Osteonics Corp.
Event
Event 98411
Status
Ongoing
Classification
Class II
Quantity
1098 units
Official record key
device-enforcement:Z-1572-2026

Official wording

Reason: A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Code information: Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;

Distribution pattern: Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.