ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.
Z-1822-2025
Product summary
- Event
- Event 96693
- Status
- Ongoing
- Classification
- Class II
- Quantity
- 1,072 packs (2,144 units) - 156 packs US, 916 packs OUS
- Official record key
device-enforcement:Z-1822-2025
Official wording
Reason: Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Code information: Siemens Material Number: 10310376. UDI-DI 00630414468198 Kit lot numbers: 86237A74, 44207A75. ***Updated 5/8/26*** Additional lots 25674A77 52363A79
Distribution pattern: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
Derived failure modes
-
Unknown
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.