Joined enrichment rows that support device recall linkage and root-cause wording.

Indexed record set

39,516 evidence rows

Page 1569 of 1581

Evidence Product Join Root cause
device-recall:cfres:114135 Medline Total Knee CDS; a single patient prescription procedure pack and disposables, including a back table cover; Product Usage: This Complete Delivery Systems (CDS) custom pack includes components necessary for a total knee replacement surgical procedure. Official recall number · Z-0532-2013 Packaging
device-recall:cfres:114620 SOL SYS L 9IN CALC 2.25/16.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Official recall number · Z-0581-2013 Device Design
device-recall:cfres:114682 BD Phoenix PMIC-108 Panels, Catalog number 448418, labeled in part ***Becton, Dickinson and Company 7 Loveton Circle Sparks, USA. The BD Phoenix" Automated Microbiology system is intended for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically significant bacteria. Official recall number · Z-0570-2013 Nonconforming Material/Component
device-recall:cfres:111792 ThyssenKrupp Accessibility LEVANT Stairlift. Intended to mechanically transport one mobility impaired person in a fold-down seat up and down straight stairs, indoors and outdoors. Official recall number · Z-0557-2013 Device Design
device-recall:cfres:114628 SOL SYS L 9IN CALC 2.25/19.5 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Official recall number · Z-0583-2013 Device Design
device-recall:cfres:108838 Medrad(R) Continuum MR Infusion System Primary VSA Stopcock Set Peristaltic Pump System designed for patients who require medications and other fluids during an MR procedure. It is intended to provide infusion therapy directly prior to, during and immediately after the MR procedure functioning while either stationary or mobile. Official recall number · Z-0599-2013 Nonconforming Material/Component
device-recall:cfres:114616 SOL SYS 8IN CALC 1.5/18.0 Packaging: Product is packed within a poly protector, with protection added as needed for fragile points. The entire stem is then placed within a foam sleeve for added protection. The assembly is then placed and sealed in an inner and outer tray for aseptic transfer. The sealed package is placed into a folding carton, labeled, and shrink wrapped. Intended Use: SOLUTION System Calcar Stems are used in total hip arthroplasty (THA) to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components. Official recall number · Z-0577-2013 Device Design
device-recall:cfres:114660 Persona, The Personalized Knee System, Tibial Broach, Cemented, Size E-F, Nonsterile, This device is used with the Persona Knee System for patients with severe knee pain and disability due to: Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. o Collagen disorders, and/or avascular necrosis of the femoral condyle. o Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy. o Moderate valgus, varus, or flexion deformities. Official recall number · Z-0548-2013 Device Design
device-recall:cfres:114418 Mindray V12 Size: 320 mm X 320 mm X 450 mm N.W.: 3 kg G.W.: 6 kg Qty:1 Manufactured in China Mindray V21 Size: 515 mm X 335 mm X 685 mm N.W.: 8 kg G.W.: 11 kg Qty:1 Manufactured in China One product, V Series monitor, available in two sizes: The V12 has a 12 inch screen, the V21 has a 21 inch screen. The V Series Monitor is a Multi Parameter Patient Monitor (with Arrhythmia Detection and Alarms) capability of performing IV Drug and Hemodynamic Calculations and interfacing with network devices. It also monitors of the following human physiological parameters: 1) ECG waveform derived from 3, 5, 6 and 12 lead measurements, 2) Heart Rate, 3) Pulse Oximetry (Sp O2), 4) ST Segment Analysis, 5) Arrhythmia Detection, 6) Non Invasive Blood Pressure (NIBP), 7) Invasive Blood Pressure (IBP), 8) Cardiac Output (CO), 9) Respiratory Gasses, 10) Respiration Rate, 11) Temperature, It is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. Official recall number · Z-0528-2013 Nonconforming Material/Component
device-recall:cfres:112107 Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body. Official recall number · Z-2256-2012 Radiation Control for Health and Safety Act
device-recall:cfres:110805 BD GeneOhm Cdiff Assay , Catalog # 441401, 200 box tests 2-3 labeled in part***GeneOhm Sciences Canada, Inc., 2555 boul, du Parc-Technologique, Quebec, QC, Canada G1P 4S5*** BD GeneOhm Cdiff Assay is a qualitative in vitro diagnostic test for the direct detection of C. difficile toxin gene (tcdB) in human liquid or soft stool specimens from patients suspected of having Clostridium difficile-associated disease (CDAD). The test, based on real-time PCR, is intended for use as an aid in diagnosis of CDAD. The test is performed directly on the specimen, utilizing polymerase chain reaction (PCR) for the amplification of specific targets and fluorogenic target-specific hybridization probes for the detection of the amplified DNA. Official recall number · Z-2319-2012 Nonconforming Material/Component
device-recall:cfres:112152 Brand name: RayStation. Generic name: Treatment Planning System. Software version and build numbers are 2.5.0.144 and 2.5.1.89. Product lifetime is 5 years. RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user. The system functionality can be configured based on user needs. The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system. Official recall number · Z-2324-2012 Device Design
device-recall:cfres:114235 Ventlab Premium Small Child Resuscitator, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028. Official recall number · Z-0177-2013 Component design/selection
device-recall:cfres:112538 Boule Con-Diff Hematology Control For in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments. Official recall number · Z-0191-2013 Nonconforming Material/Component
device-recall:cfres:113443 AquaLiner Hydrophilic Guidewire 0.035" X 260cm, Angle/ Regular. AquaLiner Hydrophilic Guidewire is designed for use in angiographic procedures to introduce/ position catheters and interventional devices within the vasculature. Official recall number · Z-0134-2013 Equipment maintenance
device-recall:cfres:112481 Battery, 100/115V (Includes charger and connecting cable) The Sarns" Centrifugal System is indicated for use in cardiopulmonary bypass procedures only. Official recall number · Z-0049-2013 Component design/selection
device-recall:cfres:112213 IDS-iSYS Intact PINP Control Set, Catalog Number IS-4030. Packaged in cardboard Box containing plastic bottles. The IDS-iSYS Intact amino-terminal propeptide of type I procollagen (Intact PINP) Control Set is used for quality control of the IDS-iSYS Intact PINP Assay on the IDS-iSYS Multi-Discipline Automated Analyser. Official recall number · Z-0045-2013 Labeling False and Misleading
device-recall:cfres:112909 Unicel Closed Tube Aliquotter (UCTA) for UniCel DxC 660i, UniCel DxC 680i, UniCel DxC 860i, UniCel DxC 880i Synchron Access Clinical Systems. Part Number A45853. Product Usage: The UniCel DxC Synchron Clinical Systems are fully automated, computer-controlled clinical chemistry analyzers designed for the in-vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent). Official recall number · Z-0059-2013 Nonconforming Material/Component
device-recall:cfres:112798 CM Flowmeter Pkg H, Model #36700, Serial # 13749 Product Usage: Conscious Sedation Unit Analgesia Gas Machine. Official recall number · Z-0019-2013 Device Design
device-recall:cfres:113036 3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for indoor use only. Official recall number · Z-0022-2013 Component design/selection
device-recall:cfres:112799 Newport Portable Package, Model #50000, Serial # 13591, 13592, 13604, 13606, 13614, 13662, 13663, 13664, 13726, 13731, 13733, 13736, 13737, 13738, 13740, 13768, 13770, 13771, 13772, 13784, 13787. Product Usage: Portable Analgesia Gas Machine. Official recall number · Z-0020-2013 Device Design
device-recall:cfres:112206 IDS-iSYS Ostase¿ BAP (Bone Alkaline Phosphatase) Control Set, Catalog Number IS-2830. Packaged in cardboard Box containing plastic bottles. The IDS-iSYS Ostase¿ BAP (IDS-iSYS BAP) Control Set is used for quality control of the IDS-iSYS Ostase¿ BAP (Bone Alkaline Phosphatase) Assay on the IDS-iSYS Multi-Discipline Automated Analyser. Official recall number · Z-0041-2013 Labeling False and Misleading
device-recall:cfres:112034 Precise Digital Accelerator Delivery of radiation to defined target volumes Official recall number · Z-2405-2012 Nonconforming Material/Component
device-recall:cfres:111524 CMS-4873 Custom CT Biopsy Tray, CMS-4873-R1 Custom CT Biopsy Tray containing 4 oz. bottle PVP Prep Solution Product Usage: Usage: The PVP contained in the kit is used in skin prep Official recall number · Z-2357-2012 Material/Component Contamination
device-recall:cfres:112270 Covidien Duet TRS 45 3.5 mm Universal Straight Single Use Loading Unit. Product Code: DUET4535. Single Use Loading Unit (SULU) tissue reinforcement stapler. The Endo GIA Universal Staplers with Duet TRS reloads have applications in abdominal, gynecologic and pediatric surgery for resection, transection and creation of anastomoses. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. Official recall number · Z-2432-2012 Device Design